p-Chlorobenzenesulfonic acid (pCBSA) is a by-product from the manufacture
of DDT.
pCBSA is determined using LC-MS/MS. Water samples are injected directly onto the LC-MS/MS. The instrument is calibrated from 5 to 500 ug/L. An MDL of 0.6 ug/L was obtained, therefore a reporting limit of 2 ug/L has been adopted. The chromatography is performed on an anion exchange column with an eluent composed of dilute potassium hydroxide in methanol - water. The eluent is passed through an anion suppressor (Dionex) prior to the MS. Ionization is accomplished using an electrospray interface (Varain 1200L LC-MS/MS). The instrument was operated in the MS/MS mode, detecting negative ions. |
The fragmentations monitored include:
For more information click on LC-MS/MS. Other applications include: Perchlorate and Warfarin. |