Home > Technical Articles > Metals Screen

Screening Samples for Metals using ICPMS

Introduction

Inductively coupled plasma-mass spectrometry (ICPMS) can be used to screen samples for metals contamination.  In a typical scan of a sample in solution, data may be  collected for masses 5-250 amu over a 

1-2 minute period with sensitivity in the sub ug/L range for most elements. This means that the concentration of ~68 elements can be determined quickly. Using this technique, WCAS offers a test which we call a "Metals Screen" to measure the concentration of ~68 elements in a sample.  An example report appears below:

~68 Elements

Sample: Acid Leachate
Metals Screen Report
Inductively Coupled Plasma - Mass Spectrometry
 

ppb

Detection Limit

 

ppb

Detection Limit

Aluminum

1660

0.7

  Molybdenum 

ND

0.1

Antimony

0.9

0.1

  Neodymium

2

0.1

Arsenic

ND

0.1

  Nickel

ND

0.3

Barium

19

0.1

  Niobium

0.5

0.1

Beryllium

0.2

0.1

  Osmium

ND

0.1

Bismuth

ND

0.1

  Palladium

ND

0.1

Boron

ND

3

  Phosphorus

ND

20

Bromine

ND

50

  Platinum

ND

0.1

Cadmium

ND

0.1

  Potassium

ND

10

Calcium

9100

10

  Praseodymium

0.4

0.1

Cerium

3

0.1

  Rhenium

ND

0.1

Cesium

ND

0.1

  Rhodium

ND

0.1

Chromium

ND

1

  Rubidium

0.8

0.1

Cobalt

ND

0.1

  Ruthenium

ND

0.1

Copper

0.8

0.2

  Samarium

0.3

0.1

Dysprosium

0.3

0.1

  Selenium

ND

1

Erbium

ND

0.1

  Silicon

ND

50

Europium

ND

0.1

  Silver

ND

0.1

Gadolinium

0.3

0.1

  Sodium

230

3

Gallium

0.6

0.1

  Strontium

18

0.1

Germanium

ND

0.1

  Tantalum

ND

0.1

Gold

ND

0.1

  Tellurium

ND

0.1

Hafnium

ND

0.1

  Thallium

0.8

0.1

Holmium

ND

0.1

  Thorium

ND

0.1

Iodine

ND

0.1

  Thulium

ND

0.1

Iridium

ND

0.1

  Tin

ND

0.4

Iron

ND

20

  Titanium

6

0.2

Lanthanum

2

0.1

  Tungsten

ND

0.1

Lead

1

0.1

  Uranium

0.3

0.1

Lithium

1

0.1

  Vanadium

0.9

0.2

Lutetium

ND

2

  Ytterbium

ND

0.1

Magnesium

1400

0.4

  Yttrium

2

0.1

Manganese

120

0.1

  Zinc

10

0.2

Mercury

ND

0.1

  Zirconium

3

0.1

Date Analyzed: 2-18-97
Elements Not Analyzed: All Gasses, C, S, Sc, In, Tb

~50 elements fully quantitated (in blue);

others estimated +50% but can also be quantitated on request

Detection Limits will depend on sample preparation

Detection Limits based on blanks

Alternative techniques provide lower detection limits: CV-AFS, preconcentration, etc.

Includes Uncommon Elements

XRF and ICP-OES also available

 

Isotope dilution techniques for lead provide very accurate results

Sc, In, Tb used as internal standards


How's it done?

In a typical test, a sample is put into solution usually by digesting or diluting in acid.  The sample solution is spiked with known concentrations of three elements (usually Sc, In, and Tb).  Standards of common elements, blanks, and the sample solution are then introduced to the ICPMS  The element concentrations in the samples are then determined one of two ways:   (1) for the elements that are  in the

 standards, the elements are determined quantitatively (within 15% for most) or (2) for the elements not in the standards, the elements are estimated (within 50%) by comparison with internal standard responses. Unlike many other spectroscopic techniques ICPMS has a fairly uniform response, so fairly accurate estimates can be made.

For more information on the analytical technique, click here.


Pros

Since each element has a unique isotope mass, interferences are usually minimal.  Since detection limits are in the sub ug/L range, even tiny samples (1-10 mg) can be analyzed with excellent detection limits.  This is a very efficient way to test expensive or rare samples like synthetic peptides or proteins, or even samples from ancient artwork. 

Only 2-5 mL of sample solution are needed.  Usually a minimum of 10 mL is requested so that duplicates, spikes, and dilutions can be analyzed.  Do you suspect metals contamination is causing a problem with your product?  This is an excellent method for solving this problem.

Some non-metallic elements are usually reported using this method: Br, I, and P.


Cons

High concentrations of elements that are predominant in the sample can cause interferences for other elements.  These are usually elements of neighboring masses, oxides, halides, argides of some elements, and doubly charged ions.  For example Ca based products have an interference for Ni from CaO, and carbon based samples have an interference for Cr from ArC.  All of these potential interferences are considered and flagged in your report.  Our systems have collision cells to minimize molecular ion interferences.

Some elements are not determined by this technique: gaseous elements, C, S, F, Cl, and the internal standard elements that are spiked (usually Sc, In, and Tb).   Some samples are very difficult to place in solution (e.g. silicates).   Alternatives such as X-ray fluorescence which do not require solutions may be considered. 

For a quotation......

For a full listing of Exova services, please visit our new Exova web site,  Exova Pharmaceutical Solutions web site and our new Santa Fe Springs laboratory home page.

Exova

9240 Santa Fe Springs Rd
Santa Fe Springs, CA 90670
562.948.2225   Fax 562.948.5850
info400@exova.com

Exova

Formerly West Coast Analytical Service (WCAS) and Bodycote Testing Group