Update 6/2007

USP has issued a bulletin pushing back the date of implementation of the General Notices statement on Residual Solvents from July 1, 2007 to July 1, 2008.

As before we still offer the new methods as well as the old Organic Volatile Impurities <467> method.  For the new method we need to know what compounds to look for as that will determine what specific method to follow for testing.

 

USP Revision Bulletin


USP <467>  Residual Solvents 4/2007

There are many different solvents which are used in pharmaceutical manufacturing. Previously this test was called Organic Volatile Impurities (OVI).  As most of you probably already know, major changes to USP <467> are on the horizon (July 1, 2007). These changes have been in the works for a few years now, in an effort to harmonize USP requirements to the European Pharmacopoeia (EP).

The first change is in the title of the Chapter itself: Residual Solvents (RS) is replacing Organic Volatile Impurities (OVI). The new RS chapter now potentially tests for 53 individual solvents, instead of the four under OVI. However, the test methodology has (in some ways) been simplified, as the Chapter now describes only headspace conditions for analysis, similar to the old OVI Method IV.

Class 1, 2 and 3 Solvents

The 53 compounds are broken into three classes. Class 1 solvents should not be employed due to high toxicity or deleterious effects on the environment, unless their use is unavoidable. Class 2 solvents are less toxic, but should still be limited. Their concentration limits range from 50-3880 ppm. Class 3 solvents are not known to present a health hazard at levels normally acceptable in pharmaceuticals (< 5000 ppm).

There is also a major change in how the testing is applied. Individual monographs will no longer list a Residual Solvent requirement. However, that does not mean that the testing is not required. It has been replaced by a statement in the General Notices, requiring that Residual Solvents be controlled in drug product.

 


Maximum Limits

This leads to one of the primary issues regarding the new requirements. The maximum limits apply ONLY to finished goods (drug products). Drug substances, along with excipients, can exceed the concentration limits given in the Chapter, although they must then be labeled to that effect. Also, the concentration limits given assume a daily dose of 10 grams. For smaller doses, limits are expressed as a Permissible Daily Exposure (PDE), given in micrograms per day. This is a function of the dosage of the drug product, including all excipients. In the final analysis, it is the PDE that is the determining factor as to whether a product meets the specifications or not. USP gives different options to determine this.

USP<467> Residual Solvents
to  be published in the second supplement to USP 30

An important ramification of this new approach is that, based on knowledge of the raw materials which go into the manufacture of a drug product, the final product may not need to be tested. If, for example, all of the raw materials meet the maximum concentration limits, and the daily dose of the drug product is 10 g or less, then the final product will meet the specification and no testing is required (assuming a dose of 10 g or less). This is referred to by USP as Option 1. Also, if the amounts of residual solvents in the drug substance and each of the excipients is known, then the amount present in the drug product can be calculated based on the formulation. If this calculation yields an amount less than the concentration limit, the product passes under Option 2. If the limit is exceeded, the product can still be analyzed to determine if the RS content was reduced during the manufacturing process to a level which would meet the specification. Manufacturers are allowed to use RS contents from Certificates of Analysis to perform these calculations.

The Analysis

The analytical approach described in the Chapter has also been changed. As stated above, only headspace conditions are given. Summarized, a sample is first screened for all of the Class 1 and Class 2 solvents that are amenable to the method (Procedure A). If no peaks are detected, the sample meets the specification. If any peaks corresponding to a solvent are detected, the sample is analyzed again using a different type of GC column (Procedure B). If no peak corresponding to the solvent tentatively identified in Procedure A is detected in Procedure B, then the sample meets the specification. However, if the presence of the solvent is confirmed by Procedure B, the sample is analyzed again in order to quantify the amount of solvent present (Procedure C). This is accomplished using the conditions from either Procedure A or Procedure B, whichever gives superior performance.

 

USP <467> Organic Volatile Impurities - OVI

For a quotation...

Also check out these related pages:

USP<467> OVI
the old version

Residual Solvents
a GCMS alternative to USP<467> OVI


 

For a full listing of Exova services, please visit our new Exova web site,  Exova Pharmaceutical Solutions web site and our new Santa Fe Springs laboratory home page.

Exova

9240 Santa Fe Springs Rd
Santa Fe Springs, CA 90670
562.948.2225   Fax 562.948.5850
info400@exova.com

Exova

Formerly West Coast Analytical Service (WCAS) and Bodycote Testing Group